Medonic Hematology M32 Maintenance
Resumen
TLDRThe Medonic M32 analyzer maintenance tutorial outlines four key maintenance procedures: daily, weekly, monthly, and six-monthly, applicable to all M32 series models. Daily maintenance includes system flushing and needle cleaning with alcohol, ensuring accurate readings by clearing debris. Weekly maintenance involves using a 2% Hypochlorite solution for a thorough cleaning. Monthly maintenance utilizes an enzymatic cleaner to digest blood residues, particularly for the Autoloader intake. The six-monthly maintenance employs a complete cleaning kit that includes three cleaners and ensures the internal components are sanitized and functioning correctly. Regular maintenance is vital for the accurate operation of the analyzer.
Para llevar
- 🛠️ Maintenance Procedures: Daily, Weekly, Monthly, 6-Monthly.
- 🔧 System Flushing: Clears debris before measurements.
- 🧽 Daily Cleaning: Inlet sipper pipes with 70% alcohol.
- 🧴 Weekly Cleaning: Use 2% Hypochlorite solution.
- 🔬 Monthly Cleaning: Enzymatic cleaner for blood residues.
- 📅 6-Month Maintenance: Complete cleaning kit required.
- ❌ No Household Cleaners: They can damage the system.
- ✅ Check Zero Values: Confirm instrument readiness after maintenance.
- ⏳ CLOT PREVENTION: 15 minutes for monthly cleaning.
- 🚨 Regular Maintenance: Ensures accurate operation.
Cronología
- 00:00:00 - 00:05:00
Munjin introduces the Medonic M32 analyzer maintenance procedures, which include daily, weekly, monthly, and 6-monthly tasks. The focus is on the M32S Autoloader model, applicable to all M32 and M-Series models. Key supplies mentioned are two main reagents (Diluent and Lyze) and the Boule Cleaning Kit, which contains three types of cleaners for regular maintenance. These preparations, along with clean tissues and a 70% alcohol solution, are essential for proper functioning and cleaning processes.
- 00:05:00 - 00:10:00
Daily maintenance procedures are outlined, starting with flushing the system using the PRIME SYSTEM function, followed by the cleaning of the inlet sipper pipe using 70% alcohol. Weekly maintenance involves using a 2% Hypochlorite Cleaner to flush the system, ensuring all values return to '0' for accurate results. This step is critical for using the analyzer in good condition while avoiding household cleaners that may damage the instruments.
- 00:10:00 - 00:19:34
The monthly maintenance includes a specific cleaning procedure for the Autoloader and intakes with respective cleaners. A 15-minute clot prevention procedure is highlighted for the Autoloader's enzymatic cleaner. Six-monthly maintenance requires a complete cleaning kit and involves emptying the system and using all three cleaners sequentially, ensuring all components are cleaned properly to maintain the analyzers' performance and probability of accurate measurements.
Mapa mental
Vídeo de preguntas y respuestas
What models are included in the M32 Series?
The M32 Series includes the Basic, MPA, Closed Tube, and Autoloader models (M32B, M32M, M32C, M32S).
What is the purpose of the Daily Maintenance?
Daily Maintenance includes flushing the system and cleaning the inlet sipper pipe to ensure accurate measurements.
What cleaning agents are used during maintenance?
The cleaning agents include Diluent, Lyze, Enzymatic Cleaner, 2% Hypochlorite Cleaner, and Detergent Cleaner.
How often should the 6-Monthly Maintenance be conducted?
The 6-Monthly Maintenance should be performed every six months.
Can household cleaners be used for maintenance?
No, commercial household cleaners contain perfumes and colorants that can damage the instrument.
What is the role of the Boule Cleaning Kit?
The Boule Cleaning Kit is used during monthly and half-yearly maintenance to clean the instrument effectively.
What steps are involved in the Weekly Maintenance?
Weekly Maintenance involves aspirating a sample of 2% Hypochlorite solution to clean the system.
What should be checked after completing the maintenance?
Check that all values such as RBC, WBC, PLT, and HGB show '0' to confirm the instrument's readiness.
How long does the CLOT PREVENTION procedure take during Monthly Maintenance?
The CLOT PREVENTION procedure takes about 15 minutes.
What should be done after using the Detergent Cleaner in the 6-Monthly Maintenance?
After using the Detergent Cleaner, empty the system to remove cleaner residues.
Ver más resúmenes de vídeos
Agility Robotics Broadens Relationship with Amazon
BIOLOGI Kelas 11 - Sistem Pertahanan Tubuh (PART 1) | GIA Academy
Podcast on “Role of AI in Transforming Businesses and Jobs”
Brutally Honest Cold Email Advice From Alex Hormozi
Mail server DNS records - setup and configuration explained
Henry VIII: Mind of a Tyrant Part Four with David Starkey
- Medonic M32
- maintenance procedures
- hematology analyzer
- cleaning procedures
- M32 Series
- Daily Maintenance
- Weekly Maintenance
- Monthly Maintenance
- 6-Monthly Maintenance
- Boule Cleaning Kit